Part One: Research Background Globally, approximately 500 million people suffer from
disabling hearing loss, accounting for about 6.8% of the world's population. Hearing loss
has emerged as a major global health issue. The cornerstone of hearing loss prevention
and protection lies in hearing level testing. Early detection and intervention are
crucial components of hearing healthcare. Currently, pure tone audiometry serves as the
gold standard for hearing level testing, requiring patients to visit a hospital's
standard soundproof room and complete the test under the guidance of a professional
audiologist. However, existing professional audiometers are bulky, consume high power,
and necessitate soundproof rooms (with background noise <30dBA), among other drawbacks.
Moreover, primary hospitals or units lack professional audiologists and soundproof rooms,
making it impossible to diagnose and treat hearing loss. Therefore, a portable hearing
testing device that does not require a soundproof room would be a boon for the diagnosis
and treatment of hearing loss patients. Currently, Wuhan Douting Technology Co., Ltd. has
produced a portable hearing testing device (E200), supporting frequencies from 125Hz to
16000Hz, covering both conventional and extended high frequencies. The frequency
tolerance, total harmonic distortion, and accuracy of hearing levels fully meet the
technical requirements of Grade 4 audiometers according to national standards. In
addition to meeting the functional and performance requirements of general audiometers,
this device offers the following technical advantages and innovations: 1. Combining
active and passive noise reduction, it achieves comprehensive noise reduction of over
29dBSPL, allowing for audiometry in environments with background noise <60dBA; 2.
Supporting extended high frequencies, it can detect potential hearing damage earlier; 3.
Supporting audiometry for rapid use by non-professionals, with automatic calibration,
allowing non-professionals to operate by following prompts. This device is not yet on the
market, and we intend to verify its reliability in assessing hearing conditions.
Part Two: Research Objectives To verify the reliability of a novel portable hearing
testing device in assessing hearing conditions.
Part Three: Research Design Paired design
Part Four: Inclusion and Exclusion Criteria
Inclusion Criteria:
Age: ≥6 years old, no gender restriction Able to cooperate in completing pure tone
audiometry
Exclusion Criteria:
Patients with acute otitis media or unremovable secretions blocking the external auditory
canal Patients with mental or psychological disorders or unable to cooperate with the
examination for other reasons
Part Five: Research Process This study will recruit volunteers from the outpatient and
inpatient departments of Xijing Hospital. Each volunteer will undergo two audiological
tests, with at least a 10-minute interval between the two tests to prevent intolerance to
the hearing tests. Each audiological test will take approximately 6-8 minutes per
subject.
Consultation and Physical Examination:
Researchers will collect general information, including name, gender, age, hearing
status, and ear disease history; Conduct a physical examination of the ears; If a
volunteer meets the inclusion criteria and voluntarily sign the informed consent form,
researchers will determine the testing order using a random number method based on your
enrollment sequence.
Conventional pure-tone audiometry:Patients undergo conventional pure tone audiometry
using the interacoustics AC40 (Denmark) with TDH39 headphones (Telephonics; Farmingdale,
NY, USA). We conducte audiometric testing in a sound-proof booth (background noise<30dBA)
and evaluate the frequency range of 125-8000 Hz. Before the audiometry, patients will be
asked to identify which ear have better hearing, and the ear with better hearing was
tested first. This study employ the Hughson-Westlake approach, known as the " up-5
down-10 technique." The pure tone sound level was initially introduced at an estimated
higher threshold and reduced by 10 dB HL after a correct response from the patient. If
the patient did not respond correctly or at all, the sound level was increased by 5 dB
HL. When there was a difference of 40 dB HL or more in air conduction thresholds between
the ears, contralateral masking was applied to correct the cross-hearing.
E200 Audiometry: Noise-canceling headphones (RAN-3500) are used for all patients during
the E200 audiometry. This is conducted in a regular room (ambient noise: 42.6-58.7 dBA,
measured using a standard sound level meter: AWA6292 Multi-function Sound Level Meter)
adjacent to a busy hallway without a sound-proof booth to simulate an ordinary
environment. The Hughson-Westlake approach and contralateral masking are applied when
conducting the E200 audiometry.