Dexamethasone Eluting Cochlear Implant: a Pivotal Study

Last updated: September 10, 2024
Sponsor: MED-EL Elektromedizinische Geräte GesmbH
Overall Status: Active - Recruiting

Phase

N/A

Condition

Hearing Aid

Treatment

Cochlear Implant

Clinical Study ID

NCT06142682
CDEX2_MED-EL_CRD_2021_01
  • Ages 18-85
  • All Genders

Study Summary

The newly developed MED-EL Cochlear Implant Mi1250 +FLEX28 DEX (CIDEXEL) incorporates the anti-inflammatory agent dexamethasone (DEX) into the electrode array. The passive elution of DEX during the post-implantation period has the purpose of counteracting the increase of the post-operative impedance induced by the insertion trauma.

The aim of this clinical investigation is to obtain confirmatory data concerning safety and performances of the investigational device in the clinical population.

Eligibility Criteria

Inclusion

Inclusion Criteria:

Minimum age of eighteen (18) years at time of enrolment.

  • Hearing loss on the ipsilateral ear (see Figure 8 for the specific audiologicalcriteria).

  • A functional auditory nerve in the ear to be implanted.

  • Subjects reporting to having used an optimally fit hearing aids for a minimum ofthree months before the decision that a cochlear implant is the preferential option.

  • Cochlea anatomy compatible with the insertion of a FLEX28 electrode array.

  • Compatibility with a soft surgery approach as per clinical practice at the site.

  • Post-lingual hearing impairment.

  • Subject fulfilling indication criteria for a CI according to the local professionalstandards, as reported by the implanting surgeon.

  • General health condition, psychological and emotional condition deemed compatiblewith the treatment and tests performed in this study and realistic expectations, asdeemed appropriate by the implanting surgeon.

  • Signed and dated informed consent before the start of any study-specific procedure.

Exclusion

Exclusion Criteria:

  • Lack of compliance with any inclusion criterion.

  • Previously having received a cochlear implant on the ear chosen for placing the IMD.

  • Evidence of ossification or any other cochlear anomaly that might prevent completeinsertion of the electrode array or might cause increased risk of infection (e.g.dysplastic cochlea), as confirmed by medical examination and/or as per CT/MRT.

  • Evidence of anatomic abnormalities that would prevent appropriate placement of thestimulator housing in the bone of the skull.

  • Known allergic reaction or intolerance to the materials used in the implant (including medical grade silicone, platinum, iridium, parylene c, DEX).

  • Known absence of cochlear development or if the cause of deafness isnon-functionality of the auditory nerve and/or the upper auditory pathway.

  • Evidence of active external or middle ear infection or history of recurrent middleear infection in the ear to be implanted.

  • Evidence of perforated tympanic membrane in the ear to be implanted.

  • Patient reporting immunosuppressive therapy or corticosteroids therapy in the 4weeks before enrolment.

  • Evidence of concomitant use of medicinal substances that, in the opinion of theinvestigator, could alter the therapeutic efficacy of dexamethasone.

  • Unwillingness or inability of the candidate to comply with all investigationalrequirements.

  • Evidence of medical contraindications to surgery of the middle and inner ear andanaesthesia.

  • Additional disabilities that would prevent or restrict participation in theaudiological and medical evaluations required of the clinical investigation.

  • Subject not capable of giving consent.

  • Individual subjected to court-ordered or government-authorized deprivation ofliberty as defined by § 27 MPDG.

  • Evidence of Pregnancy.

Study Design

Total Participants: 60
Treatment Group(s): 1
Primary Treatment: Cochlear Implant
Phase:
Study Start date:
September 02, 2024
Estimated Completion Date:
December 31, 2025

Study Description

The Cochlear Implant Mi1250 +FLEX28 DEX (CIDEXEL) is the Investigational Medical Device (IMD) under investigation. MED-EL CI systems consist of an implantable part (i.e. CIDEXEL), external components (i.e. technically compatible audio/speech processor and its accessories) and software that controls and powers the implant. The IMD CIDEXEL will be the implantable part of this new MED-EL CI System and shall only be used together with technically compatible MED-EL external components.

None of the components mentioned above incorporates, as an integral part transplants or tissues or cells of human or animal origin. No products incorporating or derived from tissues or cells of human or animal origin were used during its manufacturing.

• Primary Objective

The primary objective of this clinical investigation is to confirm the clinical performance of the CIDEXEL by comparing Speech intelligibility results measured pre-operatively to those measured at the 6-months follow-up.

For this clinical investigation to be successful, mean speech test results at the 6-months visit shall be superior to the mean speech test results obtained pre-operatively. (Monosyllables)

• Secondary Objectives

The secondary objectives of this clinical investigation are to measure:

  • Impedance Field Telemetry (IFT) values and derived measures

  • Fitting Maps reiteration appraisal

  • Electrically evoked Compound Action Potential (eCAP) and derived measures

  • Unaided audiograms obtained pre- and post-operatively

  • Hearing Preservation (HP) %

  • The improvement in generic quality of life (measured by the Health Utilities Index (HUI2&3)) pre- to the 6-months visit

  • The improvement in disease specific quality of life (measured by the Nijmegen Cochlear Implant Questionnaire (NCIQ)) pre- to the 6-months visit

  • Surgical Usability and satisfaction (Ad Hoc questionnaire for the surgeons )

  • Sentence test in noise results across the tested intervals (OlSa)

  • Speech, spatial and quality of hearing across the tested intervals (Speech, Spatial and Qualities of hearing scale (SSQ12))

  • Sound quality over time (Hearing Implant Sound Quality Index (HISQUI))

  • Wearing time (data Logging)

  • Safety Objective

The purpose of this clinical investigation is to establish and verify the clinical safety of the CIDEXEL. To confirm the safety profile of the device which will be evaluated through the analysis of adverse events during the follow-up period. For the study to be considered a success, the results of the adverse event analysis shall never cause an unbalanced risk vs. benefit assessment biased towards the former.

Connect with a study center

  • HNO-Klinik der Ruhr-Universität Bochum

    Bochum,
    Germany

    Site Not Available

  • Klinik und Poliklinik für Hals-, Nasen- und Ohrenheilkunde

    Dresden,
    Germany

    Site Not Available

  • Universitätsklinikum Frankfurt Klinik für HNO-Heilkunde

    Frankfurt,
    Germany

    Site Not Available

  • Martin-Luther-Universität Halle-Wittenberg, Universitätsklinik und Poliklinik für Hals-Nasen-OhrenHeilkunde, Kopf- und Hals-Chirurgie

    Halle, 06097
    Germany

    Active - Recruiting

  • HNO-Klinik der Medizinischen Hochschule Hannover

    Hannover,
    Germany

    Site Not Available

  • Universitätsklinik für Hals-, Nasen- und Ohrenheilkunde

    Tübingen,
    Germany

    Site Not Available

  • Klinik und Poliklinik für Hals-, Nasen- und Ohrenkrankheiten, plastische und ästhetische Operationen. Die Universitäts-HNO-Klinik Würzburg

    Würzburg,
    Germany

    Site Not Available

Not the study for you?

Let us help you find the best match. Sign up as a volunteer and receive email notifications when clinical trials are posted in the medical category of interest to you.