This is a database study of clinically indicated MRIs on patients with permanent pacemakers and implantable cardioverter defibrillators.
Ferromagnetic Implants such as PPM's and ICD's are traditionally accepted as contraindications to MRI due to safety concerns. The hypothesis is that these procedures can be safely performed when appropriate safety precautions and vetting practices are implemented.
This is a cohort study. 1000 patients with a clinical need for MRI will be included in the study at OHSU. The type of device and leads will be ascertained and a safety protocol will be strictly adhered to.
2. Background
It has been estimated that 50-75% of patients with implanted devices (pacemakers and ICDs) will develop an indication for magnetic resonance imaging (MRI) examination. Thus, there are approximately 200,000 patients annually denied MRI scans because of the presence of PPMs and ICDs in the United States. The current standard of care is not to perform these scans, and instead to utilize other (inferior) imaging, and make clinical decisions without the benefit of MRI imaging. This is the current standard of care of OHSU, where only pacemaker patients who are not pacemaker dependent can undergo MRI scans, per radiology policy. However, this leads to gaps in care for some patients who would clearly benefit from an MRI scan. In some centers, it has been shown that when proper precautions are taken, MRIs can be safely performed on patients with PPMs and ICDs. The published experience from Johns Hopkins reports over 500 scans performed without any adverse outcomes.(1) There are also several smaller studies documenting the safe experience with MRI scanning in device patients (reviewed in (1), table 4)
The main concerns which have led to restrictions on MRIs in patients with cardiac rhythm devices are the following:(2)
While all of the concerns are real and have been reported in earlier generation devices, with proper patient selection and monitoring during the scan, MRIs can be performed safely on most device patients. Johns Hopkins has an experience of over 500 scans without any adverse events.(1) The proposed protocol here follows exactly the Hopkins protocol, and, indeed, the OHSU data will be combined with the Hopkins data for future publications.
To review the most recent Johns Hopkins publication:(1) 555 MRI studies in 438 device (PPM and ICD) patients were reported. All MRI scans were clinically indicated, and a device clinician, familiar with device interrogation and programming, was physically present for the entire duration of the scan. All patients were continuously monitored via telemetry and pulse oximetry during the MRI. For pacemaker patients, if they are pacemaker dependent they were reprogrammed to a non-sensing mode for the scan. If they were not pacemaker dependent, they were left in their usual mode. For ICD patients, if they were not pacemaker dependent, they were left in their usual mode. However, pacemaker dependent ICD patients were not scanned, as ICDs generally do not have a non-sensing pacing mode. Therapy for tachycardia was disabled in all ICDs for the scan.
There were three clinical events during scans, all were "power on resets", which means that the device resets to "out of the box" original factory settings. One of these patients had an ICD, felt tugging in his chest during the scan, and the scan was not completed. The other two power on resets were in PPM patients, in whom the scan was completed. Neither were pacemaker dependent, and neither noticed anything abnormal during the scan. All three patients had normally functioning devices following the scan, and at followup.
In the overall patient population in the Hopkins study, no single patient had a malfunction of a device immediately following the scan. There was a slight change in the electrical parameters of the leads, on average, but this did not result in the need for any intervention on any patient.
3. Objectives
The primary objective of this study is to develop a protocol, consistent with DHHS/CMS guidelines and requirements to further document safety in clinically indicated MR imaging in patients with implanted cardioverter defibrillators (ICD's) and pacemakers.
4. Study Procedures
MRI's will be conducted on patients who require a clinically indicated MRI only. All studies will be done in the 1.5 Tesla MRI unit. ECG monitoring pads will be placed on the patients for the duration of the study. An external defibrillator and ACLS drugs will be on hand. Heart rate, blood pressure, O2 saturation will also be monitored non-invasively throughout the study. All devices will undergo a complete interrogation prior to imaging. Parameters such as atrial and ventricular pacing thresholds, R and P wave amplitudes, lead impedance, and battery status will be measured and recorded. PPMs will be programmed to an asynchronous mode if dependent and to an inhibited mode in patients without pacemaker dependence.
The pacemaker function of ICD's in pacemaker independent patients will be programmed to ventricular inhibited (VVI) mode at 50 bpm. Pacemaker dependent patients with ICDs will be excluded.
In addition, there is a theoretical concern that patients who have capped or abandoned leads in their chest may be at risk for heating. While there is little or no objective data to support this concern, given this theoretical concern, patients in this category will only be admitted to the study if the referring physician deems the MRI critical and after the PI has carefully considered the risk/benefit ration for each patient who falls in to this category.
Similarly, there historically has also been a concern that newly implanted leads (<4 weeks) are at risk for lead to dislodgement d/t possible torque from the MRI environment. This too is only a theoretical concern and there is little or no scientific evidence to support this claim. Nevertheless, patients with newly implanted pacemaker/ICD leads will be considered by the PI for protocol inclusion only if the referring physician deems the MRI to be critical and after considering the risk/benefit to each individual patient.
After the appropriate MRI protocol for each MRI patient's unique condition has been completed, the device will be re-programmed to its original settings and completely interrogated to detect any changes in device performance.
The duration of the study extends to the post procedure interrogation at 1-6 weeks post MRI. IF the PPM or ICD reveals any malfunction post imaging (not seen yet), the patient will be followed by the electrophysiology service and appropriate follow-up management will be arranged.
5. Study Statistics
There are no studies showing any health hazards associated with magnetic field exposure. PPM and ICD implants subjected to the strong electromagnetic field of MR imaging are thought to be susceptible to movement and torque. Other traditional concerns include malfunction due to demagnetization or magnetic activation leading to inappropriate pacing and shocks and artificial sensing of electromagnetic interference from MRI leading to inappropriate inhibition and/or activation of antitachycardia therapy, device malfunction and induced arrhythmias, as detailed in the background section above. Heating of the device and surrounding tissues leading to fibrosis and increased capture thresholds have also been a theoretical concern. However, recent work has revealed the safety of PPM's and certain ICD's in the MR environment given appropriate precautions(1, 3, 4).
2. Steps taken to minimize the risks. Appropriate measures discussed in the protocol section will be taken to minimize risks associated with device exposure to MRI. Consistent with standard MRI policy, patients in the first trimester of pregnancy and patients with history of allergic reactions to gadolinium will undergo imaging without any contrast media.
3. Plan for reporting unanticipated problems or study deviations. Patient data will be monitored on a per patient basis. Pre-vs. post MRI device testing data will be compared and any significant differences immediately reported to the IRB.
7. Benefits a. Description of the probable benefits for the participant and for society. The patient will benefit from a standardized protocol to minimize the risk of an MR scan deemed necessary for his/her care which would otherwise be denied or performed without any organized framework. Society will benefit because of the increasing number of implanted devices and expanding use of MRI.
8 Costs
The patient and his/her insurance will be responsible for the cost of the clinically indicated MRI.
Condition | MRI |
---|---|
Treatment | MRI |
Clinical Study Identifier | NCT03915015 |
Sponsor | Oregon Health and Science University |
Last Modified on | 2 February 2022 |
,
You have contacted , on
Your message has been sent to the study team at ,
You are contacting
Primary Contact
Additional screening procedures may be conducted by the study team before you can be confirmed eligible to participate.
Learn moreIf you are confirmed eligible after full screening, you will be required to understand and sign the informed consent if you decide to enroll in the study. Once enrolled you may be asked to make scheduled visits over a period of time.
Learn moreComplete your scheduled study participation activities and then you are done. You may receive summary of study results if provided by the sponsor.
Learn moreEvery year hundreds of thousands of volunteers step forward to participate in research. Sign up as a volunteer and receive email notifications when clinical trials are posted in the medical category of interest to you.
Sign up as volunteer
Lorem ipsum dolor sit amet consectetur, adipisicing elit. Ipsa vel nobis alias. Quae eveniet velit voluptate quo doloribus maxime et dicta in sequi, corporis quod. Ea, dolor eius? Dolore, vel!
No annotations made yet
Congrats! You have your own personal workspace now.